- Conventional Pollution Online Monitoring Instrument
- Drinking Water Online Monitoring Instrument
- Special Parameter Online Monitoring Instrument
- Heavy Metal Online Monitoring Instrument
- Industrial Process Water Online Monitoring Instrument
- Anodic Stripping Voltammetry Heavy Metal Monitoring Instrument
- Laboratory Online Testing Instrument
- Water Treatment Process Instruments
System Overview
COD-NH3N-8000-COD ammonia nitrogen 2-in-1 online analyzer is a new generation of online analyzer in line with domestic standards and industry standards. The instrument and the realization of an instrument can simultaneously determine the chemical oxygen demand (COD) in water and ammonia nitrogen in water, and the online determination of the two pollution factors are in line with the current national standards and industry requirements, saving the user 50% of the use cost.
Potassium dichromate was used for high temperature digestion and colorimetric determination. The analytical process and results were in line with the national standards and industry requirements of GB11914-89 and HJ/T399-2007. The on-line ammonia nitrogen analyzer adopts the method of Knoll reagent or salicylic acid optical colorimetric determination. The analysis process and analysis results meet the national standards and industry requirements of HJ536-2009 and HJ/T101-2003.
System Features
· Simultaneous determination of chemical oxygen demand (COD) and ammonia nitrogen content in water;
· Two-in-one monitoring of COD and ammonia nitrogen, saving 50% of the use cost;
· Flexible calibration of standard solution to ensure high accuracy;
· Unique optical quantitative algorithm ensures high repeatability;
· The flexible setting of reaction time ensures that any water sample can be accurately monitored;
· Long service life of the injection pump, only need to replace the syringe once a year;
· The external sampling adopts two ways: submersible pump and self-priming pump;
· Online monitoring methods are diversified, which can realize manual measurement at any time, automatic timing measurement, automatic periodic measurement and other measurement methods.
· Advanced self-diagnosis, alarm system, when the sample suddenly interrupted, there is a prompt signal.
· Simple operation and maintenance.
Technical Parameters
Measurement methods: | chromium method, Knoll reagent method or salicylic acid method; |
Measuring range: | COD: 0-1000mg /L;
ammonia nitrogen: 0-50mg/L; |
Accuracy: | <10%; |
Repeatability: | <5%; |
Response time (>90%) : | about 30min; |
Measurement method: | timing, equal interval, manual; |
Correction mode: | automatic timing correction or manual correction; |
Sampling inside the instrument: | using injection pump; |
External sampling: | submersible pump and self-priming pump are provided; |
Data transmission: | 4-20mA; RS232; RS485; |
Display: | 8.0 “large screen LCD display, resolution of 800×600; |
Reagent change: | 3~5 weeks; |
Data storage: | one year of valid data; |
Operating temperature: | +0 ~ +40°C; |
Power supply: | 220 VAC±10%/50-60Hz; |
Power consumption: | about 100W; |
Size: | 590*700*360mm; |
Weight: | about 35kg. |