TEL:+86 0571-28993551
Conventional Pollution Online Monitoring Instrument
Total Phosphorus Online Analyzer
product details

Product model: TP-8000

 

System Overview

 

Polyphosphates and other phosphorous compounds in water are hydrolyzed under acidic conditions of high temperature and high pressure to produce phosphoric acid. For other refractory phosphorus compounds, they are oxidized to phosphate by the strong oxidizing agent sodium persulfate. In the acidic medium, orthophosphate reacts with ammonium molybdate and potassium antimony tartrate to produce phosphomolybdate heteropoly compound, which is reduced to blue phosphomolybdate by ascorbic acid, and the total phosphorus content in the water sample is obtained by measuring absorbance. In the case of phosphate measurement, phosphate in water can be directly measured without oxidation.

 

System Features

 

1.The water sample pretreatment device adopts maintenance-free design, which can ensure that the maintenance cycle of the pretreatment device exceeds half a year.

2.The chemical digestion time can be adjusted, and the determination process and results meet the relevant standards.

3.The analytical flow path design and reagent formulation of all imported devices ensure the measurement reproducibility, which can reach 5% at present.

4.Automatic operation, no personnel on duty, can achieve automatic zero, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, automatic recovery and other intelligent functions.

5.TP-8000 total phosphorus online analyzer online monitoring methods are diversified, can achieve manual measurement at any time, automatic timing measurement, automatic periodic measurement and other measurement methods.

6.Automatic leakage alarm function, when there is a reagent leak, the instrument automatically alarms, prompting the user to maintain.

 

Technical Parameters

 

Test method: ammonium phosphomolybdate colorimetric method;
Test range: 0 ~ 2/5/10/30mg /L;
Detection limit: 0.1mg/L;
Accuracy: conventional: < 10%; < 1mg/L: < ±0.5mg/L;
Repeatability: routine: < 5%; < 1mg/L: < ±0.1mg/L;
Response time (90%) : can be adjusted according to the water sample, at least 30min;
Test mode: timing, equal interval, manual;
Reagent consumption: 1mL at a time;
Maintenance mode: self-maintenance, user maintenance interval >5 months;
Self-monitoring: self-monitoring of leaks; Instrument status self-diagnosis;
Analog output: 4– 20mA analog output;
Relay control: 2 24V 1A relay high and low point control;
Data transmission mode: RS232, RS485;
Display: 8.0-inch LCD touch screen, resolution 800×600;
Data storage: one year of valid data;
Reagent change: change once a month;
Operating temperature: +0 ~ 50°C;
Power supply: 220 ±10% VAC; 50-60Hz;
Power consumption: about 80W;
Size: 530×1340×390mm;
Weight: about 45Kg.