Product Model: T8000-LA

 

Measurement Principle

 

Concentration colorimetric determination by online extraction technology, which is a proprietary technology of Moody Technology, reduces the detection limit of lanthanum through this technology, and reduces the detection limit to 0.002ppm through online extraction technology department.

 

Measurement Method

 

The concentration colorimetric method is used to adjust the PH value of the water sample after masking the interfering substances. Finally, the content of lanthanum in industrial wastewater is determined by online extraction and color development after enrichment.

Multiple Options Are Available

Timing measurement enables the instrument to be started every day at any time the user wants to monitor;

Isochronous measurement can automatically start the instrument for measurement every few hours at a fixed time interval every day;

Continuous measurement can realize automatic sample measurement one by one, which can be used for product acceptance and related technical certification;

Manual measurement can realize the user to start the measurement at any time, and can be used for field experiment comparison and equipment installation and debugging.

 

Technical Parameter

 

Measuring range: (0-0.1/0.5/1) mg/L lanthanum;
Measurement accuracy: <10%;
Repeatability: <5%;
Zero drift: ±0.005mg/L;
Range drift: ±10%;
MTBF (trouble-free running time) : ≥720 h/ time;
Actual water sample comparison: ±10%;
Measurement time: can be arbitrarily set, generally 15min;
Correction mode: automatic timing correction or manual correction;
External sampling: submersible pump and self-priming pump are provided respectively;
Data transmission: can provide 4-20MA, RS232, etc.;
Ambient temperature: +5°C to +40°C;
Mechanical size: 530×1340×390mm;
Weight: about 45kg.
Power supply: (220±20) VAC /(50±0.5) Hz;
Power consumption: about 100W.

 

Benefits

 

Maintenance Free Design

Water sample pretreatment device adopts the design of free maintenance, to ensure preprocessing device maintenance cycle more than half a year.

 

Chemical Digestion Time Can Be Adjusted

The chemical digestion time can adjust the measurement process and results to meet the relevant standards of the environmental protection industry and water quality analysis industry.

 

High Reproducibility

The fully imported components, innovative analytical flow design, and reagent formulation ensure high measurement reproducibility, with a current measurement reproducibility of up to 5%.

 

Fully Automatic Operation

No need for personnel on duty, it can achieve intelligent functions such as automatic zero adjustment, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, and automatic recovery.

 

Diversified Online Monitoring Methods

It can achieve measurement methods such as manual measurement at any time, automatic timed measurement, and automatic periodic measurement.

 

Automatic Liquid Leakage Alarm Function

When there is a reagent leak, the instrument will automatically alarm and prompt the user to perform maintenance.

 

No Need for Professional Training

Operation and maintenance are extremely simple, without the need for professional training.