Product Model: T8000-Mo

 

Typical Application

 

Automatic analysis and monitoring of molybdenum content in domestic sewage and industrial wastewater; Continuous automatic monitoring of molybdenum in industrial circulating water.

 

Measurement Principle

 

The water samples were injected into the digestion tank by injection pump, and then the reducing agent was injected to reduce the various forms of molybdenum in water to the same valence state. Secondly, the buffer solution is injected again, adjusted to the appropriate PH value, and then the characteristic color developer is added, molybdenum and the color developer in the water produce orange-yellow complex, and then the photoelectric colorimeter is used to measure the color change of the orange-yellow complex at a certain wavelength, and then the content of molybdenum in the water is calculated according to the Lambert-Beer law.

 

Instrument Features

 

· The water sample pretreatment device adopts maintenance-free design, which can ensure that the maintenance cycle of the pretreatment device exceeds half a year;

· The reaction time can be adjusted to reduce the time required to measure molybdenum in water;

· All imported components and innovative analytical flow path design and reagent formulation ensure high measurement reproducibility, currently up to 5%;

· Automatic operation, no personnel on duty, can achieve automatic zero, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, automatic recovery and other intelligent functions;

· On-line monitoring methods are diversified, which can realize manual measurement at any time, automatic timing measurement, automatic periodic measurement and other measurement methods;

· Fast response speed, the test takes about 10 minutes;

· Automatic leakage warning function, when there is a reagent leak, the instrument automatically warns, prompting the user to maintain;

· The test results are highlighted with large digital data display;

 

Technical Parameters

 

Measurement method: optical colorimetry;
Measuring range: 0-0.1/0.5/1.0/5/10mg/L;
Reproducibility: better than 5%;
Accuracy: ± 10%;
Measurement cycle: about 10 minutes;
Instrument calibration: automatic;
Reagent replacement cycle: 3 months;
Display: 7-inch color LCD touch screen, resolution 800×600;
Self-test system: self-monitoring of leaks; Instrument status self-diagnosis;
Analog output: 2 channels 4– 20mA analog output;
Relay control: 2 24V 1A relay high and low point control;
Data communication: RS232;
Data storage: 3,000 groups;
Operating temperature: +5 ~ +50°C;
Power supply: 220 ±10% VAC; 50-60Hz;
Power consumption: about 50 VA;
Size: 530×1340×390mm.