TEL:+86 0571-28993551
Heavy Metal Online Monitoring Instrument
Total Manganese Online Analyzer
product details

Product Model: T8000-Mn

 

System Overview

 

T8000-Mn Total Manganese online analyzer is a new generation of automatic online monitoring equipment based on the standard method of China’s environmental protection industry.

 

Principle

 

After the pre-treated water sample is injected into a special reactor by an injection pump, it first reacts with a reducing reagent to uniformly reduce all forms of manganese in the water sample to bivalent manganese ions, then adjusts the PH value of the solution, and finally adds a characteristic color developer for color reaction. Within the measurement range, the degree of color change is proportional to the total manganese concentration in the water sample. By measuring the degree of color change, the amount of total manganese in the water sample can be calculated.

 

System Features

 

· The use of optical quantitative technology can ensure the high accuracy of measurement;

· The water sample pretreatment device adopts maintenance-free design, which can ensure that the maintenance cycle of the pretreatment device exceeds half a year;

· The measurement process and results can meet the relevant standards;

· Microsampling technology ensures low consumption of reagents;

· All imported devices, analysis flow path design and reagent formulation ensure the measurement reproducibility, the current measurement reproducibility can reach 5%;

· Automatic operation, no personnel on duty, can achieve automatic zero, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, automatic recovery and other intelligent functions;

· The total manganese online analyzer has a variety of online monitoring methods, which can realize manual measurement at any time, automatic timing measurement, automatic periodic measurement and other measurement methods.

 

Technical Parameters

 

Test method: photoelectric colorimetry;
Test range: 0-0.5/1/5/10/20mg/L;
Accuracy: <10%;
Reproducibility: <5%;
Response time (>90%) : about 10min;
Test method: timing measurement, equal interval measurement, manual measurement at any time;
Correction mode: automatic timing correction;
Pretreatment maintenance: automatic recoil cleaning;
Daily maintenance: automatic maintenance, user maintenance interval >1 month;
Self-test system: self-monitoring of leaks; Instrument status self-diagnosis;
Analog output: 4-20mA analog output;
Relay control: 2 way 24V1A relay high and low point control;
Data transmission mode: RS232, RS485;
Display: 8.0 “large screen color LCD display, touch screen, resolution 800*600;
Data storage: one year of valid data;
Operating temperature: +5 ~ 40°C;
Power supply: 220±10%VAC; 50-60Hz;
Power consumption: about 50VA;
Size: 530×1340×390mm;
Weight: about 45Kg.