- Conventional Pollution Online Monitoring Instrument
- Drinking Water Online Monitoring Instrument
- Special Parameter Online Monitoring Instrument
- Heavy Metal Online Monitoring Instrument
- Industrial Process Water Online Monitoring Instrument
- Anodic Stripping Voltammetry Heavy Metal Monitoring Instrument
- Laboratory Online Testing Instrument
- Water Treatment Process Instruments
Product Model: T8000-As
Measurement Principle
After the sample is filtered, it is pumped into the reactor, and the reducing agent is injected first to reduce the arsenic pentavalent to arsenic trivalent. Then the gas generator is added to convert the measured arsenic into a gas and be expelled, and the expelled arsenic gas is absorbed by the solution in the colorimetric tank. After absorption, the arsenic gas reacts with the absorption solution. The analyzer calculates the absorbance of the solution according to the color of the solution, and then calculates the total arsenic concentration in the sample according to its stored correction factor.
Notable Features
Stable and Reliable
In order to meet the requirements of industrial and environmental online, the electrical part and the chemical pipeline part are completely isolated, and the simple and stable optical detection device ensures that the instrument can run steadily for a long time.
Easy to Install
The T8000-As water arsenic online analyzer has successfully passed a series of tests to facilitate installation and setting, and only needs to connect the reagent, sample, waste liquid pipeline and power cord during installation, and can be started after setting parameters.
Automatic Calibration
The instrument terminates the analysis and performs calibration according to the calibration time selected by the user, checks and stores new calibration parameters.
Measurement Interval
The user can set the analyzer to remain in standby mode between two measurements to avoid drug waste.
Technical Parameters
Measurement principle: | optical colorimetry; |
Measurement interval: | can be set arbitrarily; |
Measurement time: | about 30 minutes; |
Measuring range: | 0-0.5/1/5ppm, other ranges can be extended; |
Range expansion: | with online dilution function, can meet the requirements of high, medium and low range; |
Minimum detection limit: | 0.001mg/L; |
Reproducibility: | 5%; |
Signal output: | standard 4-20mA analog output, maximum load 400 ohms or 0-5V, other RS485 or RS232 optional; |
Signal input: | 1 channel analysis, 1 channel correction; |
Alarm: | 1 way high limit alarm, 1 way correction; |
Sample and waste liquid transport: | no pressure; Sample temperature: 10-30℃; |
Agent replacement: | 3~4 weeks according to the operating temperature changes; |
Ambient temperature: | 0-40 ℃; |
Protection class: | IP55; |
Power supply: | 220VAC; |
Weight: | 45kg (excluding chemicals); |
Size: | 530×1340×390mm. |
Advantages
· Fully automatic operation;
· Long time control, low maintenance, low operating cost;
· Low drug consumption, short preparation time;
· Simple maintenance, no special electrician training required;
· Isolation of electrical and chemical parts;
· Intelligent control with embedded control system, fully automatic operation;
· Using large-screen color LCD touch screen to display monitoring results and human-machine dialogue, it can store valid data for at least one year;
· With self-diagnostic function, can identify whether water sample is missing;
· Standard 4-20mA analog output;
· At the same time, RS232 digital interface is provided, which can be connected with local or remote PC through the corresponding software to achieve remote control;
· After power failure, it has the function of self-starting the call.