TEL:+86 0571-28993551
Industrial Process Water Online Monitoring Instrument
Water Quality Alkalinity Online Analyzer
product details

Product model: TOH900

 

Measurement Principle

 

The instrument uses the continuous flow analysis technology to complete the automatic sample sampling and automatic analysis, through the automatic titration technology to measure the optical signal changes in the mixing process of the sample and titrant to complete the automatic alkalinity analysis, and finally the total alkalinity value of the actual water sample is calculated by comparing with the standard known alkalinity material.

 

System overview

 

The TOH900 online water quality alkalinity analyzer is typically used to measure the total alkalinity of industrial boiler water, circulating water and drinking water.

 

Measurement method

 

optical titration.Can Achieve a Variety of Choices:

Timing measurement can be realized every day in any user want to monitor the time to start the instrument for measurement;

Isochronous measurement can automatically start the instrument for measurement every few hours at a fixed time interval every day;

Continuous measurement can realize automatic sample measurement one by one, which can be used for product acceptance and related technical certification;

Manual measurement can realize the user to start the measurement at any time, and can be used for field experiment comparison and equipment installation and debugging;

 

Technical Parameters

 

Measuring range: 0 ~ 500mg/L CaCO3;
Accuracy: ± 3%;
Reproducibility: ± 1%;
Response time: 90% response, less than 10 minutes;
Measurement time: 8 minutes;
Test mode: timing, equal interval, manual;
Maintenance mode: self-maintenance, user maintenance interval >5 months;
Self-monitoring: self-diagnosis of instrument status;
Analog output: 4– 20mA analog output;
Relay control: 2 24V 1A relay high and low point control;
Data transmission mode: RS232, RS485;
Display: 8.0-inch LCD touch screen, resolution 800×600;
Data storage: one year of valid data;
Operating temperature: +0 ~ 40°C;
Power supply: 220 ±10% VAC; 50-60Hz;
Power consumption: about 100 VA;
Size: 590×700×360mm;
Weight: about 35KG.

 

System Features

 

· Can realize automatic unattended online real-time monitoring of water quality alkalinity;

· The water sample pretreatment device adopts maintenance-free design, which can ensure that the maintenance cycle of the pretreatment device exceeds half a year;

· The measurement process and results can meet the relevant industry standards;

· Microsampling technology ensures low consumption of reagents;

· All imported devices, analysis flow path design and reagent formulation ensure measurement reproducibility, which can reach 1% at present;

· Automatic operation, no personnel on duty, can achieve automatic zero, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, automatic recovery and other intelligent functions;

· Diversified online analysis methods can realize manual measurement at any time, automatic timing measurement, automatic periodic measurement and other measurement methods.