Product model: An-8000

 

System Overview

 

Modi AN-8000 aniline in water online analyzer is An automatic online monitoring equipment controlled by an embedded system, which can be applied to a variety of water quality such as river water, surface water, dye industry wastewater, pharmaceutical industry wastewater and chemical industry wastewater online real-time water quality monitoring. After the sample is filtered, it is pumped into the reactor, and the interfering substances are eliminated by decolorization and masking, and then the pH value of the solution is adjusted to make the solution have a suitable pH, and then the characteristic color developer is added to react with aniline in water for color development, and the absorbance of the reactants is measured. The concentration of aniline in the sample was calculated by the absorbance value and the correction equation stored in the monitor.

 

System Features

 

· The water sample pretreatment device adopts maintenance-free design, which can ensure that the maintenance cycle of the pretreatment device exceeds half a year.

· The chemical reaction time can be adjusted, and the determination process and results meet the domestic standards.

· All imported devices, analysis flow path design and reagent formulation ensure measurement reproducibility, which can reach 5% at present.

· Automatic operation, no personnel on duty, can achieve automatic zero, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, automatic recovery and other intelligent functions.

· Online monitoring methods are diversified, which can realize manual measurement at any time, automatic timing measurement, automatic periodic measurement and other measurement methods.

· Automatic leakage warning function, when there is a reagent leak, the instrument automatically warns, prompting the user to maintain.

· Operation and maintenance is extremely simple, no professional training is required

· The electrical part and the fluid part are isolated, and the embedded control system is used for fully automatic operation.

 

Technical Parameters

 

Measurement principle: optical colorimetry;
Range: 0 ~ 0.5/2.0/5.0mg/L;
Measurement type: automatic timing measurement or manual measurement at any time;
Measurement interval: can be set arbitrarily;
Measurement time: about 20 minutes;
Measurement accuracy: 10%;
Minimum detection limit: 0.05mg/L;
Reproducibility: 5%;
Signal output: standard 4-20mA analog output, maximum load 400 ohms or 0-5V RS232 digital output (optional);
Sample and waste liquid transport: no pressure; Sample temperature: 0-50℃;
Agent replacement: 3~4 weeks according to the operating temperature changes;
Ambient temperature: 5-40 ℃;
Protection class: IP55;
Power: power supply: 220VAC;
Weight: 35kg (excluding chemicals);
Size: 590×700×360mm.