TEL:+86 0571-28993551
Industrial Process Water Online Monitoring Instrument
Online Water Acetone Oxime Analyzer
product details

Product Model: DMKO-8000

 

Product Overview

 

DMKO-8000 online water acetone oxime analyzer is an automatic online monitoring equipment controlled by embedded system, which can be applied to real-time online monitoring of acetone oxime content in various water quality.

 

Principle

 

After the sample is filtered, it is pumped into the reactor for the digestion reaction of the water sample, and then the pH value of the solution is adjusted to make the solution have the appropriate pH, and then the characteristic color developer is added for the color reaction, and the absorbance of the reactant is measured. The concentration of acetone oxime in the sample was calculated by the absorbance value and the correction factor stored in the monitor.

 

System Features

 

· The reagent uses a special storage device, which can extend the replacement cycle of the reagent.

· The time of chemical digestion can be adjusted, and the determination process and results meet the relevant standards in China.

· All imported devices, analysis flow path design and reagent formulation ensure measurement reproducibility, which can reach 5% at present.

· Automatic operation, no personnel on duty, can achieve automatic zero, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, automatic recovery and other intelligent functions.

· Online monitoring methods are diversified, which can realize manual measurement at any time, automatic timing measurement, automatic periodic measurement and other measurement methods.

· Operation and maintenance is extremely simple, no professional training is required.

 

Technical Parameters

 

Measurement principle: photoelectric colorimetry;
Range: 0 ~ 100/500 ppb;
Measurement type: cyclic measurement;
Measurement interval: can be set arbitrarily;
Measurement time: about 10 minutes;
Measurement accuracy: 10%;
Minimum detection limit: 5ppb;
Reproducibility: 5%;
Signal output: standard 4-20mA analog output, maximum load 400 ohms or 0-5V, other RS485 or RS232 optional;
Agent replacement: 1~2 weeks according to the operating temperature changes;
Ambient temperature: 0-30 ℃;
Power supply: 220VAC;
Weight: 35kg (excluding chemicals);
Size: 590×700×360mm.