TEL:+86 0571-28993551
Special Parameter Online Monitoring Instrument
Phenol Online Analyzer  
product details

Product model: T8000-ph

 

Principle

 

4-aminoantipyrine photoelectric colorimetric method.

 

Measurement Method

 

· The content of perchlorate (perchlorate) in surface water and industrial wastewater was measured by color development after enrichment.

· can achieve a variety of choices, timing measurement can be realized every day in any user want to monitor the time to start the instrument for measurement;

· Isochronous measurement can automatically start the instrument for measurement every few hours at a fixed time interval every day;

· Continuous measurement can realize automatic sample measurement one by one, which can be used for product acceptance and related technical certification;

· Manual measurement can realize the user to start the measurement at any time, and can be used for field experiment comparison and equipment installation and debugging;

 

System description

 

The T8000-ph online phenol analyzer in water is a fully automatic online monitoring instrument controlled by an embedded system, which can be applied to various water quality such as river water, surface water, and industrial wastewater. After the sample is filtered, it is pumped into the reactor. First, a masking agent is injected to eliminate the interfering substances. Then, the pH value of the solution is adjusted to make it have a suitable pH value. Then, a characteristic chromogenic agent is added to react with phenols in water, and the absorbance of the reactant is measured; Calculate the concentration of phenol in the sample through the absorbance value and the correction factor stored by the monitor.

 

Specification

 

Range: 0-1 (mg/L;
Measurement type: cyclic measurement;
Measurement interval: can be set arbitrarily;
Measurement time: approximately 30 minutes;
Measurement accuracy: 10%;
Minimum detection limit: 0.01mg/l;
Reproducibility: 5%;
Signal output: standard 4-20mA analog output, maximum load 400 ohms or 0-5V, other RS485 or RS232 options;
Signal input: 1 channel analysis, 1 channel correction;
Sample and waste liquid transportation: no pressure; Sample temperature: 10-30 ℃;
Reagent replacement: 3-4 weeks depending on the operating temperature;
Environmental temperature: 5-40 ℃;
Protection level: IP55;
Power supply: 220VAC;
Weight: 45kg;
Size: 530x1340x390mm.

 

Characteristics

 

· The water sample pre-treatment device adopts a maintenance free design, which can ensure that the maintenance cycle of the pre-treatment device exceeds six months.

· The chemical digestion time can be adjusted, and the determination process and results meet the national standard HJ503-2009.

· The design of fully imported devices and analysis flow paths, as well as the formulation of reagents, ensure extremely high measurement reproducibility.Currently, the measurement reproducibility can reach 5%.

· Fully automatic operation, without the need for personnel on duty, can achieve intelligent functions such as automatic zero adjustment, automatic calibration, automatic measurement, automatic cleaning, automatic maintenance, self-protection, and automatic recovery.

· Diversified online monitoring methods can achieve measurement methods such as manual measurement at any time, automatic timing measurement,and automatic periodic measurement.

· Automatic liquid leakage alarm function, when there is a reagent leakage, the instrument will automatically alarm and prompt the user for maintenance.

· Operation and maintenance are extremely simple, without the need for professional training.

· The electrical and fluid parts are isolated, and an embedded control system is used for fully automatic operation.